ICP vs XRF Assay: Which Precious Metal Testing Should You Trust?
If you sell precious metal scrap, the assay is the invoice. One number — the percentage of metal in your material — decides whether you are paid fairly or quietly underpaid. XRF and ICP are the two methods buyers use to produce that number, and they are not interchangeable. Here is what each one actually does, where each fails, and which to demand before you sign a settlement.
The Short Answer
XRF is a screening tool. ICP is a settlement tool. XRF tells you quickly and non-destructively what your material probably contains, which makes it ideal for sorting incoming lots, identifying alloys, and catching obvious errors. ICP destroys a representative sample, dissolves it, and measures the metal content with far greater accuracy — which is why serious refiners use ICP (or classical fire assay) to price a shipment, and why a buyer who offers only a handheld XRF reading is asking you to accept a rough estimate as final.
How XRF Works — and Where It Stops Being Useful
X-ray fluorescence fires X-rays at a sample and reads the characteristic energy emitted back. It is fast (seconds), non-destructive, and requires almost no sample preparation. For mixed scrap coming through a warehouse door, that is genuinely valuable.
Its limits matter just as much:
- It only reads the surface. XRF penetration is measured in microns to a few millimetres depending on the alloy. Plating, oxide layers, contamination, or a clean-looking outer skin can all mislead the reading.
- It assumes the sample is homogeneous. A crucible, a spent catalyst, or a batch of powder is usually not. Point the gun at three places and you can get three different answers.
- Spectral overlaps are real. Adjacent platinum group metals have overlapping peaks — platinum and iridium among them — so results depend heavily on the instrument’s calibration library and the operator’s competence.
- It is semi-quantitative for trace and minor elements, which is exactly the range that often decides the value of a PGM lot.
None of this makes XRF a bad instrument. It makes it a screening instrument. Using it as the final word on a settlement is like pricing a house from a photograph.
How ICP Works — and Why Refiners Rely On It
Inductively coupled plasma (ICP-OES or ICP-MS) works the other way round. A small, carefully taken sample is dissolved in acid, nebulised into a plasma at roughly 6,000–10,000 °C, and the emitted light is measured to determine concentration — parts per million or better, across many elements in a single run.
The trade-offs:
- It is destructive. The aliquot you submit is consumed. That is why reputable refiners keep a split sample and documentation.
- It takes hours, not seconds. Dissolution of iridium, rhodium, and ruthenium can be slow and requires specialised chemistry.
- Its accuracy depends entirely on the sample. This is the single most important point on this page: an ICP result is only as good as the sample that went into the tube. If the sample was not representative, a perfectly executed analysis will produce a perfectly wrong number.
Side-by-Side Comparison
Factor XRF ICP (OES / MS) Best used for Sorting, alloy ID, quick screening Settlement pricing, precise content Accuracy Semi-quantitative, surface-based High, bulk-content based Sample Non-destructive Destructive (split sample retained) Speed Seconds Hours to days Homogeneity needed Assumed (often wrong) Managed through sampling protocol Typical role Gate check before a lot is accepted Referee method for payment The Real Risk Is Not the Instrument — It’s the Sample
Most disputes we see are not about whether the lab ran the machine correctly. They are about whether the 5 grams that went into the machine represented the 500 kilograms that were shipped.
Good sampling practice looks like this:
- Homogenise first. Melt, mill, or at minimum crush and cone-and-quarter the lot so that any aliquot is representative.
- Take more than one sample. Duplicate or triplicate aliquots expose disagreement early.
- Retain a split. Both parties should hold sealed, labelled material from the same lot, so a disputed result can be re-tested.
- Agree on an umpire in advance. Name an independent laboratory and state in the contract that its result is binding if the two parties’ assays differ beyond an agreed tolerance.
These four clauses cost nothing to negotiate before shipping and are nearly impossible to fix afterwards.
What This Means When You Sell Scrap
A practical sequence for any PGM lot:
- Ask how the lot will be assayed before you agree to a price. “We will test it” is not an answer; “ICP on a homogenised sample, with a retained split” is.
- Treat a pure-XRF quote as indicative only. It is a reasonable starting point for a small lot; it is not a basis for settling kilograms of iridium or rhodium.
- Get the assay certificate in writing, with method, elements reported, and the sample identity traceable to your lot.
- Compare net returns using the same assay basis, so you are comparing prices rather than testing philosophies — our guide to getting the best price for iridium scrap walks through that comparison step by step.
How Litop Metals Tests Incoming Material
Our laboratory provides free ICP and XRF sample testing for prospective suppliers. Every lot is screened by XRF on arrival, then homogenised and assayed by ICP before a final settlement figure is issued, and a split sample is retained for verification. You receive the assay data in writing — not a verbal estimate. Whether your material is precious metal scrap in any form — crucibles, targets, thermocouple wire, powder, or spent catalyst — the same protocol applies.
- Want your material tested before you commit to a price? Send photos and approximate weights and we will arrange free XRF and ICP testing with a written result: contact our refining team.